Rekord LOINC: 57657-9
Tylko wersja angielska ACTIVE
Szczegóły rekordu LOINC
Informacje podstawowe
| Kod LOINC: |
57657-9
FHIR
|
| Long Common Name: | Disconnect sensitivity setting Ventilator alarm |
| Short Name: | Disconnect sensitivity VFr Vent alarm |
| Status: | ACTIVE |
| Class Type: | 2 |
Fully Specified Name (FSN)
EN
Oryginalna wersja angielska
Disconnect sensitivity setting
:
VFr
:
Pt
:
Ventilator.alarm
:
Qn
:
-
Klasyfikacja
| Class: |
ENPULM
|
| Consumer Name: | - |
| Order/Observation: | Observation |
| Panel Type: | - |
Informacje dodatkowe
| Units Required: | N |
| Example Units: | % |
| UCUM Units: | % |
| Version Last Changed: | 2.70 |
Definicja i opis
EN
From the Puritan Bennett 840 Operator's and Technical Reference Manual: The DSENS setting defines the percentage of returned volume lost, above which the ventilator declares a CIRCUIT DISCONNECT alarm. When DSENS is set to its lowest value (20%), it has the highest sensitivity for detecting a disconnect or leak. When DSENS is set to its highest value (95%), the ventilator has the least sensitivity for detecting a circuit disconnection, as greater than 95% of the returned volume must be lost before the alarm occurs.
Powiązane nazwy
Lung; Percent; Point in time; PULMONARY; Pulmonology; QNT; Quan; Quant; Quantitative; Random; Respiratory; Vent alarm; Volfr; Volume fraction